LAISHA
Open Laisha

FSSAI Food Analyst and Junior Analyst Examinations Quality Assurance, Statistical Quality Control & Laboratory Management: practice questions

FSSAI Food Analyst and Junior Analyst Examinations · India · Food Safety and Standards Authority of India

Original practice questions written to the published syllabus · not past papers
10worked here, free
30written to this exam in this subject
3syllabus units

About FSSAI Food Analyst and Junior Analyst Examinations

FSSAI Food Analyst and Junior Analyst Examinations is a licensing and registration exam run by Food Safety and Standards Authority of India, for candidates in India.

Official site: fssai.gov.in.

We have not been able to confirm these details against the exam body's own specification, so fees, dates and pass marks are left out on purpose. Check the current format, eligibility, fees and dates on the official site.

Quality Assurance, Statistical Quality Control & Laboratory Management in FSSAI Food Analyst and Junior Analyst Examinations

Quality systems, statistical process control, sampling inspection and running an accredited dairy laboratory.

Our 30 FSSAI Food Analyst and Junior Analyst Examinations questions in this subject: 6 easy, 17 medium, 7 hard, across 3 units.

10 practice questions, worked

Choose an answer, then open the working. Each explanation says why the right answer is right and why the others are not.

These 10 were written to FSSAI Food Analyst and Junior Analyst Examinations and are not shown on any other exam's page.

Question 1 · easy

How many standard sections does a GHS safety data sheet contain?

  1. 16
  2. 20
  3. 8
  4. 12
Answer and working

A. 16

The GHS safety data sheet has sixteen sections, from identification and hazard identification through handling, exposure controls and disposal to regulatory and other information.

Question 2 · medium

When preparing a dilute acid in the laboratory, the safe practice is to:

  1. Add the acid slowly to the water with stirring
  2. Mix equal volumes of acid and water at once
  3. Heat the acid before adding any water
  4. Add the water quickly to the acid to dissolve it
Answer and working

A. Add the acid slowly to the water with stirring

Adding acid to water lets the large volume of water absorb the heat of dilution. Adding water to concentrated sulphuric acid can boil locally and spatter acid.

Question 3 · hard

Why do many milk-recording and payment laboratories avoid potassium dichromate as a preservative for composite samples?

  1. It changes the fat content measured by mid-infrared analysers by half
  2. It is toxic and carcinogenic, so safer preservatives are preferred
  3. It is banned for all laboratory use in every country
  4. It causes rapid growth of lactic acid bacteria in the sample
Answer and working

B. It is toxic and carcinogenic, so safer preservatives are preferred

Hexavalent chromium compounds such as potassium dichromate are toxic and carcinogenic, and their disposal is problematic, so many laboratories use less hazardous preservatives where the method allows. It is not universally banned; it suppresses rather than promotes bacterial growth.

Question 4 · hard

A plant laboratory is asked to culture and confirm Brucella from raw milk samples. What is the most appropriate response?

  1. Run it only after adding potassium dichromate to the samples
  2. Run it on the open bench beside routine total plate counts
  3. Send the work to a laboratory with the higher containment it requires
  4. Run it in the chemistry laboratory, where fume cupboards are already fitted
Answer and working

C. Send the work to a laboratory with the higher containment it requires

Brucella is a well-known cause of laboratory-acquired infection, and work with its cultures needs higher biosafety containment than routine indicator counts. Fume cupboards protect against chemical vapours, not aerosols of pathogens; preservatives do not make culture safe.

Question 5 · medium

Under ISO/IEC 17025:2017, what is 'option B' for the management system?

  1. Having the management system audited by ISO itself in Geneva
  2. Using only the technical clauses and omitting internal audit and management review
  3. Seeking accreditation for calibration rather than for testing
  4. Operating an ISO 9001 system that also supports the laboratory requirements
Answer and working

D. Operating an ISO 9001 system that also supports the laboratory requirements

Clause 8 lets a laboratory meet the management-system requirements either directly (option A) or through a system run to ISO 9001 that supports the technical requirements (option B). Internal audit is required either way, and ISO does not audit laboratories.

Question 6 · medium

In measurement uncertainty, what is a Type B evaluation?

  1. Evaluation by a second laboratory in an interlaboratory study
  2. Evaluation of uncertainty only for biological tests
  3. Evaluation by means other than statistics of repeated measurements
  4. Evaluation from the standard deviation of repeated measurements
Answer and working

C. Evaluation by means other than statistics of repeated measurements

Type A components are evaluated statistically from repeated observations; Type B components come from other information such as calibration certificates, tolerances and published data. The labels say nothing about the type of test.

Question 7 · hard

Using the convention LOD = 3 × s(blank) and LOQ = 10 × s(blank), what are the LOD and LOQ when the standard deviation of blank results is 0.4 µg/kg?

  1. LOD 0.8, LOQ 1.2 µg/kg
  2. LOD 1.2, LOQ 4.0 µg/kg
  3. LOD 4.0, LOQ 1.2 µg/kg
  4. LOD 0.4, LOQ 1.2 µg/kg
Answer and working

B. LOD 1.2, LOQ 4.0 µg/kg

LOD = 3 × 0.4 = 1.2 µg/kg; LOQ = 10 × 0.4 = 4.0 µg/kg. Other conventions exist, which is why the stem states this one. Swapping them or using 1× and 3× gives the wrong pairs.

Question 8 · medium

Organisations that run proficiency-testing schemes are accredited against which standard?

  1. ISO/IEC 17043
  2. ISO/IEC 17025
  3. ISO 9001
  4. ISO 19011
Answer and working

A. ISO/IEC 17043

PT providers are accredited to ISO/IEC 17043. ISO/IEC 17025 is for testing and calibration laboratories, ISO 9001 for quality management systems and ISO 19011 gives guidance on auditing.

Question 9 · medium

What is the difference between a correction and a corrective action?

  1. They are the same thing under ISO/IEC 17025
  2. A correction fixes the instance; corrective action removes the cause
  3. A correction is done by auditors; corrective action by customers
  4. Corrective action is optional unless the customer complains
Answer and working

B. A correction fixes the instance; corrective action removes the cause

Correction deals with the immediate nonconformity (for example repeating a test); corrective action finds and eliminates the root cause so it does not recur, and its effectiveness is checked.

Question 10 · medium

Where no proficiency scheme exists for a test, how can a laboratory provide external evidence of its validity?

  1. Using only results from new analysts trained that month
  2. Repeating its own tests until the results agree
  3. Interlaboratory comparison or analysis of certified reference materials
  4. Asking the client to accept the results without evidence
Answer and working

C. Interlaboratory comparison or analysis of certified reference materials

Interlaboratory comparisons with a competent laboratory, and analysis of certified reference materials (produced under ISO 17034), give external evidence when no PT scheme is available. Repeating one's own tests only shows repeatability.

A study plan for Quality Assurance, Statistical Quality Control & Laboratory Management in FSSAI Food Analyst and Junior Analyst Examinations

This plan follows where our questions actually fall, so the time goes where the marks are likely to be. It is our reading of the syllabus, not the exam body's weighting.

  1. First, the heavy units (37% of the 30 questions): ISO/IEC 17025 and laboratory accreditation (11, 37%).

    Read the manual chapters for these first and do every worked question you can find on them.

  2. Then the middle band: Laboratory set-up, safety and chemical management (10, 33%).

    Aim for one sitting per unit, then mixed questions across them.

  3. Then the lighter units: Laboratory audits and proficiency testing (9, 30%).

    Learn the definitions, classifications and standard numbers; these units are more often recalled than worked.

  4. Leave time for the hard ones: ISO/IEC 17025 and laboratory accreditation (3 hard of 11), Laboratory audits and proficiency testing (2 hard of 9), Laboratory set-up, safety and chemical management (2 hard of 10).

    Come back to these after a first pass, untimed, and read every explanation, including for the options you rejected.

  5. Bank the easy marks: Laboratory set-up, safety and chemical management (4 easy of 10); ISO/IEC 17025 and laboratory accreditation (2 easy of 11).

    Do these early in revision so they are fast on the day.

  6. Finish timed. Sit the subject as a timed set without notes, mark it, and send each wrong answer back to its unit above.

Last updated · How Laisha writes and checks its pages · Report a mistake

Ask Laisha about this

Opens Laisha's chat with your question ready to send. Sign in or start free.

Unlock every question — FSSAI Food Analyst and Junior Analyst Examinations Quality Assurance, Statistical Quality

US$10 a month · one price worldwide · cancel any time

  • Every practice question in this subject, with full working
  • Timed exam papers and progress by unit
  • Ask Laisha to explain any answer — up to 1,000 questions a month
Free or Smarter
FreeSmarter
Grow guides, pest pages, articlesRead allRead all
Ask Laisha5 a dayUp to 1,000 a month
Exam practiceThe sample on each pageEvery question, timed papers
Crop calendar and pest alerts—For your own fields
Calculators and tools—All
Uploads and reports—100 a month, real files
10 days free for every new account, no card.