LAISHA
Open Laisha

FSSAI Food Analyst Examination Food Quality Assurance, Statistical Process Control & Laboratory Accreditation: practice questions

FSSAI Food Analyst Examination (and Junior Analyst Examination) · India · Food Safety and Standards Authority of India

Original practice questions written to the published syllabus · not past papers
10worked here, free
39written to this exam in this subject
8syllabus units

About FSSAI Food Analyst Examination

FSSAI Food Analyst Examination (and Junior Analyst Examination) is a licensing and registration exam run by Food Safety and Standards Authority of India, for candidates in India. A computer-based multiple-choice test in either the Chemical or the Biological discipline (+4 for a correct answer, -1 for a wrong one, 50% to qualify), then a practical examination, document verification and two weeks of compulsory training (11th Food Analyst Examination, FSSAI).

Source: sites.fssai.gov.in.

Checked against the exam body's published information. Formats and fees change; confirm the current edition there before you book.

Food Quality Assurance, Statistical Process Control & Laboratory Accreditation in FSSAI Food Analyst Examination

Controlling quality in the plant and in the laboratory: quality systems and TQM, control charts and process capability, acceptance sampling and two- and three-class sampling plans, laboratory quality management and ISO/IEC 17025 accreditation, method validation and proficiency testing.

Our 39 FSSAI Food Analyst Examination questions in this subject: 10 easy, 27 medium, 2 hard, across 8 units.

10 practice questions, worked

Choose an answer, then open the working. Each explanation says why the right answer is right and why the others are not.

These 10 were written to FSSAI Food Analyst Examination and are not shown on any other exam's page.

Question 1 · medium

For the same plan (n = 5, c = 2, m = 10², M = 10³ cfu/g), one unit gives 2000 cfu/g and the others are below 100. What is the decision?

  1. Accept: the average is below M
  2. Reject: one unit exceeds M
  3. Retest the high unit and use the lower value
  4. Accept: only one unit exceeds m
Answer and working

B. Reject: one unit exceeds M

Any unit above M makes the lot unacceptable regardless of c. Averaging or retesting to replace a result is not permitted by the plan.

Question 2 · medium

Why is chain-of-custody documentation essential for official food samples?

  1. It records the price paid for the sample by the officer
  2. It replaces the need for method validation in the laboratory
  3. It proves who held the sample and when
  4. It shows the sample was analysed in duplicate by two analysts
Answer and working

C. It proves who held the sample and when

Legal samples must be shown to be the sample taken, unaltered; custody records track every transfer. Price, validation and duplicates are separate matters.

Question 3 · easy

What is the purpose of a reagent blank in a chemical analysis batch?

  1. To check the repeatability of duplicates
  2. To calibrate the balance used for weighing
  3. To detect contamination or interference from reagents
  4. To measure recovery of a spiked analyte
Answer and working

C. To detect contamination or interference from reagents

A blank carried through the procedure without sample shows contribution from reagents, glassware and environment. Recovery uses spikes; balances are calibrated with weights; duplicates check precision.

Question 4 · medium

What reproducibility RSD does the Horwitz equation predict for an analyte at 1 mg/kg (mass fraction 10⁻⁶)?

  1. 4%
  2. 16%
  3. 8%
  4. 45%
Answer and working

B. 16%

RSD_R = 2^(1 − 0.5 log₁₀C) = 2^(1 + 3) = 16%. 4% is the prediction at 1% (C = 0.01); 45% at 1 µg/kg; 8% at about 0.1%.

Question 5 · medium

A 100 mL volumetric flask has a stated tolerance of ±0.08 mL with no further information. What is the standard uncertainty from the tolerance?

  1. 0.033 mL
  2. 0.080 mL
  3. 0.046 mL
  4. 0.040 mL
Answer and working

C. 0.046 mL

With a rectangular distribution, u = a/√3 = 0.08/1.732 = 0.046 mL. 0.080 uses the half-width directly; 0.040 divides by 2; 0.033 divides by √6 (triangular).

Question 6 · medium

A contaminant has a legal limit of 10 µg/kg. A laboratory reports 11.5 ± 2.0 µg/kg (k = 2). Under the rule "non-compliant if result − U exceeds the limit", what is the decision?

  1. Not shown to be non-compliant
  2. Non-compliant, since 11.5 exceeds 10
  3. Compliant beyond doubt, since 11.5 − 2.0 is below 10
  4. Retest until a result below 10 is obtained
Answer and working

A. Not shown to be non-compliant

11.5 − 2.0 = 9.5, which is not above 10, so non-compliance is not established beyond reasonable doubt under this rule; but neither is compliance proved, since 11.5 + 2.0 exceeds the limit. Retesting until a convenient result appears is not valid.

Question 7 · medium

A mass is obtained by difference of two weighings with standard uncertainties 0.03 g and 0.04 g. What is the standard uncertainty of the difference?

  1. 0.07 g
  2. 0.05 g
  3. 0.0025 g
  4. 0.01 g
Answer and working

B. 0.05 g

For sums and differences, absolute uncertainties combine in quadrature: √(0.03² + 0.04²) = 0.05 g. 0.01 subtracts them; 0.07 adds them; 0.0025 is the variance, not its square root.

Question 8 · medium

To which standard are producers of reference materials accredited?

  1. ISO 17034
  2. ISO 19011
  3. ISO/IEC 17043
  4. ISO/IEC 17020
Answer and working

A. ISO 17034

ISO 17034 covers reference-material producers; ISO/IEC 17043 covers proficiency-testing providers; ISO/IEC 17020 inspection bodies; ISO 19011 is audit guidance.

Question 9 · hard

Which Codex guideline recommends accreditation, proficiency testing, validated methods and internal QC for laboratories involved in import and export control of foods?

  1. CXG 27-1997
  2. CXC 1-1969
  3. CXG 50-2004
  4. CXG 21-1997
Answer and working

A. CXG 27-1997

CXG 27-1997 covers assessment of the competence of testing laboratories involved in import and export control. CXG 50-2004 is on sampling, CXG 21-1997 on microbiological criteria and CXC 1-1969 on food hygiene.

Question 10 · medium

How does accreditation differ from certification?

  1. Accreditation is voluntary; certification is always compulsory
  2. Accreditation: competence; certification: conformity
  3. Certification is done by accreditation bodies only
  4. They are the same process for laboratories
Answer and working

B. Accreditation: competence; certification: conformity

Accreditation (by a body under ISO/IEC 17011) recognises competence of conformity-assessment bodies such as labs; certification (by a certification body) attests that a system, product or person conforms to requirements. Either may be voluntary or required depending on the context.

A study plan for Food Quality Assurance, Statistical Process Control & Laboratory Accreditation in FSSAI Food Analyst Examination

This plan follows where our questions actually fall, so the time goes where the marks are likely to be. It is our reading of the syllabus, not the exam body's weighting.

  1. First, the heavy units (56% of the 39 questions): Method validation: accuracy, precision, LOD, LOQ and recovery (8, 21%); Laboratory accreditation bodies and their role (7, 18%); Laboratory quality management and ISO/IEC 17025 (7, 18%).

    Read the manual chapters for these first and do every worked question you can find on them.

  2. Then the middle band: Measurement uncertainty (7, 18%); Proficiency testing and reference materials (5, 13%); Two- and three-class microbiological sampling plans (3, 8%).

    Aim for one sitting per unit, then mixed questions across them.

  3. Then the lighter units: Raw-material, in-process and finished-product quality control (1, 3%); Statistical process control: control charts for variables and attributes (1, 3%).

    Learn the definitions, classifications and standard numbers; these units are more often recalled than worked.

  4. Leave time for the hard ones: Statistical process control: control charts for variables and attributes (1 hard of 1), Laboratory accreditation bodies and their role (1 hard of 7).

    Come back to these after a first pass, untimed, and read every explanation, including for the options you rejected.

  5. Bank the easy marks: Proficiency testing and reference materials (2 easy of 5); Two- and three-class microbiological sampling plans (1 easy of 3).

    Do these early in revision so they are fast on the day.

  6. Finish timed. Sit the subject as a timed set without notes, mark it, and send each wrong answer back to its unit above.

Last updated · How Laisha writes and checks its pages · Report a mistake

Ask Laisha about this

Opens Laisha's chat with your question ready to send. Sign in or start free.

Unlock every question — FSSAI Food Analyst Examination Food Quality Assurance, Statistical Process Control & Labor

US$10 a month · one price worldwide · cancel any time

  • Every practice question in this subject, with full working
  • Timed exam papers and progress by unit
  • Ask Laisha to explain any answer — up to 1,000 questions a month
Free or Smarter
FreeSmarter
Grow guides, pest pages, articlesRead allRead all
Ask Laisha5 a dayUp to 1,000 a month
Exam practiceThe sample on each pageEvery question, timed papers
Crop calendar and pest alerts—For your own fields
Calculators and tools—All
Uploads and reports—100 a month, real files
10 days free for every new account, no card.