Central and State Pollution Control Board recruitment Environmental Monitoring & Instrumental Analysis: practice questions

Central and State Pollution Control Board recruitment (Scientist, Environmental Engineer, AEE) · India · Central Pollution Control Board and State Pollution Control Boards / state PSCs

Original practice questions written to the published syllabus · not past papers
10worked here, free
23written to this exam in this subject
9syllabus units

About Central and State Pollution Control Board recruitment

Central and State Pollution Control Board recruitment (Scientist, Environmental Engineer, AEE) is a government recruitment exam run by Central Pollution Control Board and State Pollution Control Boards / state PSCs, for candidates in India. Varies by board: written objective or descriptive test in environmental science or engineering, sometimes with interview.

Official site: cpcb.nic.in.

We have not been able to confirm these details against the exam body's own specification, so fees, dates and pass marks are left out on purpose. Check the current format, eligibility, fees and dates on the official site.

Environmental Monitoring & Instrumental Analysis in Central and State Pollution Control Board recruitment

How environmental samples are taken and analysed and how good the numbers are: sampling plans, gravimetric and volumetric methods, spectroscopy, chromatography, quality assurance and monitoring networks.

Our 23 Central and State Pollution Control Board recruitment questions in this subject: 6 easy, 15 medium, 2 hard, across 9 units.

10 practice questions, worked

Choose an answer, then open the working. Each explanation says why the right answer is right and why the others are not.

These 10 were written to Central and State Pollution Control Board recruitment and are not shown on any other exam's page.

Question 1 · medium

Within what time should a cooled sample be analysed for BOD under typical standard method guidance?

  1. 28 days
  2. 48 hours
  3. 2 minutes
  4. 6 months
Answer and working

B. 48 hours

BOD changes as microbes act; 48 hours is the usual holding time; 6 months applies to acidified metals.

Question 2 · hard

What is isokinetic sampling in stack monitoring?

  1. Sampling at constant temperature to avoid condensation
  2. Sampling at the same time every day for comparability
  3. Sampling at the stack gas velocity
  4. Sampling with equal volumes from several stacks
Answer and working

C. Sampling at the stack gas velocity

Faster or slower sampling would under- or over-collect large particles because of their inertia.

Question 3 · medium

A filter weighs 1.2210 g before and 1.2345 g after filtering 250 mL of water and drying at 103–105 °C. What is the TSS?

  1. 13.5 mg/L
  2. 5.4 mg/L
  3. 540 mg/L
  4. 54 mg/L
Answer and working

D. 54 mg/L

Mass = 13.5 mg; TSS = 13.5 × 1000/250 = 54 mg/L. 13.5 forgets volume; 5.4 and 540 are factor-of-ten slips.

Question 4 · medium

Which AAS technique is used to measure mercury?

  1. Graphite furnace without reduction
  2. Cold vapour AAS
  3. Hydride generation for arsenic only
  4. Flame AAS with air–acetylene
Answer and working

B. Cold vapour AAS

Mercury is reduced to elemental vapour at room temperature and measured by cold vapour AAS.

Question 5 · medium

Which electrode is a standard method for measuring fluoride in drinking water?

  1. Platinum ORP electrode
  2. Glass pH electrode
  3. Lanthanum fluoride ISE
  4. Clark oxygen electrode
Answer and working

C. Lanthanum fluoride ISE

The LaF3 crystal ISE responds to fluoride; TISAB fixes ionic strength and releases complexed fluoride.

Question 6 · medium

A sample contains 8.4 µg/L. After adding 10 µg/L, it measures 18.2 µg/L. What is the spike recovery?

  1. 84 per cent
  2. 182 per cent
  3. 46 per cent
  4. 98 per cent
Answer and working

D. 98 per cent

(18.2 − 8.4)/10 × 100 = 98%. 182 uses the spiked result alone; 84 the original; 46 divides by the spiked result.

Question 7 · easy

Which QC sample detects contamination introduced during field handling and transport?

  1. Calibration standard
  2. Field blank
  3. Certified reference material
  4. Matrix spike
Answer and working

B. Field blank

Field blanks are analyte-free water taken through field procedures; spikes test recovery; CRMs test accuracy.

Question 8 · easy

Which international standard specifies requirements for the competence of testing laboratories?

  1. ISO 50001
  2. ISO/IEC 17025
  3. ISO 14001
  4. ISO 9001 only
Answer and working

B. ISO/IEC 17025

ISO/IEC 17025 is for laboratory competence; 14001 environmental management; 50001 energy; 9001 general quality.

Question 9 · medium

In India, which body accredits testing laboratories to ISO/IEC 17025?

  1. National Green Tribunal
  2. Central Ground Water Board
  3. NABL
  4. Bureau of Energy Efficiency
Answer and working

C. NABL

NABL accredits laboratories; MoEFCC separately recognises environmental labs under the EP Act.

Question 10 · easy

Which principle do continuous analysers commonly use to measure NO and NO2?

  1. Non-dispersive infrared
  2. Gravimetry
  3. Beta attenuation
  4. Chemiluminescence
Answer and working

D. Chemiluminescence

NO reacts with ozone emitting light; BAM is for PM; NDIR for CO.

A study plan for Environmental Monitoring & Instrumental Analysis in Central and State Pollution Control Board recruitment

This plan follows where our questions actually fall, so the time goes where the marks are likely to be. It is our reading of the syllabus, not the exam body's weighting.

  1. First, the heavy units (52% of the 23 questions): Sampling and preservation of water, soil, air and biota (5, 22%); Quality assurance, quality control and laboratory accreditation (4, 17%); Continuous and low-cost sensor monitoring (3, 13%).

    Read the manual chapters for these first and do every worked question you can find on them.

  2. Then the middle band: Gravimetric and volumetric analysis (3, 13%); Monitoring objectives and network design (3, 13%); Atomic absorption and ICP techniques (2, 9%).

    Aim for one sitting per unit, then mixed questions across them.

  3. Then the lighter units: Detection limits, accuracy, precision and calibration (1, 4%); Electrochemical methods and sensors (1, 4%); UV-visible spectrophotometry and colorimetry (1, 4%).

    Learn the definitions, classifications and standard numbers; these units are more often recalled than worked.

  4. Leave time for the hard ones: Quality assurance, quality control and laboratory accreditation (1 hard of 4), Sampling and preservation of water, soil, air and biota (1 hard of 5).

    Come back to these after a first pass, untimed, and read every explanation, including for the options you rejected.

  5. Bank the easy marks: Continuous and low-cost sensor monitoring (2 easy of 3); Quality assurance, quality control and laboratory accreditation (2 easy of 4).

    Do these early in revision so they are fast on the day.

  6. Finish timed. Sit the subject as a timed set without notes, mark it, and send each wrong answer back to its unit above.

Last updated · How Laisha writes and checks its pages · Report a mistake

Ask Laisha about this

Opens Laisha's chat with your question ready to send. Sign in or start free.

Unlock every question — Central and State Pollution Control Board recruitment Environmental Monitoring & Instrumen

US$10 a month · one price worldwide · cancel any time

  • Every practice question in this subject, with full working
  • Timed exam papers and progress by unit
  • Ask Laisha to explain any answer — up to 1,000 questions a month
Free or Smarter
FreeSmarter
Grow guides, pest pages, articlesRead allRead all
Ask Laisha5 a dayUp to 1,000 a month
Exam practiceThe sample on each pageEvery question, timed papers
Crop calendar and pest alerts—For your own fields
Calculators and tools—All
Uploads and reports—100 a month, real files
10 days free for every new account, no card.