ICAR AICE-JRF/SRF Dairy Analysis & Instrumental Methods: practice questions

ICAR AICE-JRF/SRF (PhD): Dairy Chemistry, Dairy Microbiology, Dairy Technology, Dairy Engineering · India · NTA for ICAR

Original practice questions written to the published syllabus · not past papers
10worked here, free
89written to this exam in this subject
13syllabus units

About ICAR AICE-JRF/SRF

ICAR AICE-JRF/SRF (PhD): Dairy Chemistry, Dairy Microbiology, Dairy Technology, Dairy Engineering is a PhD, fellowship and lectureship exam run by NTA for ICAR, for candidates in India.

Source: icar.org.in.

Checked against the exam body's published information. Formats and fees change; confirm the current edition there before you book.

Dairy Analysis & Instrumental Methods in ICAR AICE-JRF/SRF

Adulteration and analytical methods, and instrumental analysis: principles of spectroscopy, chromatography, electrophoresis and immunoassays, infrared analysers and their calibration, and method validation.

Testing milk and dairy products: platform tests, reference methods, adulterant detection and modern instruments.

Our 89 ICAR AICE-JRF/SRF questions in this subject: 17 easy, 45 medium, 27 hard, across 13 units.

10 practice questions, worked

Choose an answer, then open the working. Each explanation says why the right answer is right and why the others are not.

These 10 were written to ICAR AICE-JRF/SRF and are not shown on any other exam's page.

Question 1 · hard

Milk from a herd with high mastitis passes the clot-on-boiling test but flocculates in the alcohol test. Why?

  1. The COB test is more sensitive to acidity than the alcohol test at every level
  2. Alcohol stability falls with abnormal milk before COB fails
  3. Mastitic milk contains alcohol-producing yeasts
  4. Boiling removes mastitis bacteria, so the COB test cannot detect them
Answer and working

B. Alcohol stability falls with abnormal milk before COB fails

Alcohol stability falls with salt and protein changes in abnormal milk before the milk coagulates on boiling.

Question 2 · hard

What is the purpose of intercorrection factors in infrared milk analysers?

  1. To adjust each reading for the breed and lactation stage of the cow that gave the milk
  2. To convert micrometres into nanometres
  3. To correct for room humidity
  4. To correct each component's reading for the influence of the other components
Answer and working

D. To correct each component's reading for the influence of the other components

Fat absorbs slightly at the protein wavelength and so on; intercorrection removes these cross-effects.

Question 3 · hard

Why are quantitative methods and limits needed to prove addition of urea to milk?

  1. Urea evaporates during testing
  2. Milk contains urea naturally, so its presence alone proves nothing
  3. Urea is identical to lactose in all chemical tests, so it cannot be told apart from it
  4. Urea is detected only by the Gerber test
Answer and working

B. Milk contains urea naturally, so its presence alone proves nothing

Natural urea varies with feeding; only levels above a limit indicate addition.

Question 4 · hard

Which method can detect cow milk added to buffalo milk?

  1. An immunological test with antiserum against cow milk proteins
  2. The Gerber fat test
  3. The clot-on-boiling test, since cow milk clots sooner than buffalo milk
  4. The lactometer test
Answer and working

A. An immunological test with antiserum against cow milk proteins

Species-specific antibodies, electrophoresis or DNA tests identify cow milk.

Question 5 · hard

Why do animal body fats such as tallow escape detection by the phytosterol acetate test?

  1. They dissolve the sterol acetates completely
  2. They have a much higher Reichert-Meissl value than genuine ghee
  3. They contain more phytosterols than vegetable oils
  4. They contain cholesterol, not phytosterols, like milk fat
Answer and working

D. They contain cholesterol, not phytosterols, like milk fat

Body fats lack phytosterols; they lower RM and saponification values and need GC or other tests.

Question 6 · medium

A calibration line has slope 0.05 absorbance units per milligram per litre and intercept 0.01. A sample reads 0.36. What is its concentration?

  1. 7.0 milligrams per litre
  2. 0.018 milligrams per litre
  3. 7.2 milligrams per litre
  4. 18 milligrams per litre
Answer and working

A. 7.0 milligrams per litre

c = (0.36 - 0.01)/0.05 = 7.0 mg/L. 7.2 ignores the intercept.

Question 7 · hard

Why is lanthanum added when calcium in milk is determined by flame atomic absorption?

  1. To dissolve the fat
  2. To overcome phosphate interference that lowers the calcium signal
  3. To raise the flame temperature
  4. To provide the characteristic light source needed for calcium in the flame
Answer and working

B. To overcome phosphate interference that lowers the calcium signal

Phosphate forms refractory calcium compounds; lanthanum releases calcium.

Question 8 · medium

The EU reference method for cow milk in sheep, goat or buffalo cheese uses:

  1. The Reichert-Meissl value of the fat extracted from the cheese
  2. Gerber fat analysis
  3. Isoelectric focusing of gamma-caseins after plasmin digestion
  4. SDS-PAGE of whey proteins
Answer and working

C. Isoelectric focusing of gamma-caseins after plasmin digestion

IEF of gamma-caseins distinguishes bovine from other species.

Question 9 · hard

With a standard deviation of 0.004 and a calibration slope of 0.08 per unit, what is the limit of detection using 3.3 s/slope?

  1. About 3.3 units
  2. About 0.5 unit
  3. About 0.013 unit
  4. About 0.17 unit
Answer and working

D. About 0.17 unit

LOD = 3.3 x 0.004/0.08 = 0.165. 0.5 is the LOQ (10 s/slope).

Question 10 · medium

Repeatability refers to precision measured under which conditions?

  1. Same analyst, equipment and laboratory over a short time
  2. Different samples from the same batch
  3. Different methods on the same sample
  4. Different laboratories with different analysts
Answer and working

A. Same analyst, equipment and laboratory over a short time

Reproducibility involves different laboratories.

A study plan for Dairy Analysis & Instrumental Methods in ICAR AICE-JRF/SRF

This plan follows where our questions actually fall, so the time goes where the marks are likely to be. It is our reading of the syllabus, not the exam body's weighting.

  1. First, the heavy units (47% of the 89 questions): Fat determination: Gerber, Babcock and Rose-Gottlieb (9, 10%); Protein: Kjeldahl and dye-binding; total solids and SNF (9, 10%); Spectroscopy: UV-visible, infrared, atomic absorption and flame photometry (9, 10%); Method validation, reference materials and proficiency testing (8, 9%); Aflatoxin M1 detection (7, 8%).

    Read the manual chapters for these first and do every worked question you can find on them.

  2. Then the middle band: Antibiotic, pesticide and heavy-metal residue testing (7, 8%); Chromatography: TLC, GC and HPLC (7, 8%); Electrophoresis and immunoassays: SDS-PAGE, ELISA and lateral-flow tests (7, 8%); Detection of adulterants: water, starch, urea, neutralisers, detergent, vegetable fat (6, 7%); Ghee and butter fat adulteration: BR reading, Reichert-Meissl and Polenske values, sterol tests (6, 7%).

    Aim for one sitting per unit, then mixed questions across them.

  3. Then the lighter units: Infrared milk analysers and their calibration (6, 7%); Platform tests: organoleptic, COB, alcohol, sediment, lactometer (5, 6%); Sampling of milk and dairy products (3, 3%).

    Learn the definitions, classifications and standard numbers; these units are more often recalled than worked.

  4. Leave time for the hard ones: Detection of adulterants: water, starch, urea, neutralisers, detergent, vegetable fat (4 hard of 6), Ghee and butter fat adulteration: BR reading, Reichert-Meissl and Polenske values, sterol tests (3 hard of 6), Aflatoxin M1 detection (3 hard of 7).

    Come back to these after a first pass, untimed, and read every explanation, including for the options you rejected.

  5. Bank the easy marks: Platform tests: organoleptic, COB, alcohol, sediment, lactometer (3 easy of 5); Sampling of milk and dairy products (1 easy of 3).

    Do these early in revision so they are fast on the day.

  6. Finish timed. Sit the subject as a timed set without notes, mark it, and send each wrong answer back to its unit above.

Last updated · How Laisha writes and checks its pages · Report a mistake

Ask Laisha about this

Opens Laisha's chat with your question ready to send. Sign in or start free.

Unlock every question — ICAR AICE-JRF/SRF Dairy Analysis & Instrumental Methods

US$10 a month · one price worldwide · cancel any time

  • Every practice question in this subject, with full working
  • Timed exam papers and progress by unit
  • Ask Laisha to explain any answer — up to 1,000 questions a month
Free or Smarter
FreeSmarter
Grow guides, pest pages, articlesRead allRead all
Ask Laisha5 a dayUp to 1,000 a month
Exam practiceThe sample on each pageEvery question, timed papers
Crop calendar and pest alerts—For your own fields
Calculators and tools—All
Uploads and reports—100 a month, real files
10 days free for every new account, no card.