ICAR AIEEA-PG / PGS Food Analysis: Proximate, Chemical & Instrumental Methods: practice questions

ICAR AIEEA-PG / PGS — Code 18 Food Science Technology · India · ICAR through the National Testing Agency

Original practice questions written to the published syllabus · not past papers
10worked here, free
65written to this exam in this subject
14syllabus units

About ICAR AIEEA-PG / PGS

ICAR AIEEA-PG / PGS — Code 18 Food Science Technology is a postgraduate admission exam run by ICAR through the National Testing Agency, for candidates in India. Computer-based, multiple choice, +4/−1 marking.

Source: icar.org.in.

Checked against the exam body's published information. Formats and fees change; confirm the current edition there before you book.

Food Analysis: Proximate, Chemical & Instrumental Methods in ICAR AIEEA-PG / PGS

Unit I names modern analytical techniques in food analysis. MCQs test principles and applications: Beer–Lambert law, which detector for which analyte, Kjeldahl stages and factors, Gerber versus Mojonnier, crude versus dietary fibre, SDS-PAGE separation basis, sandwich versus competitive ELISA.

How a food laboratory measures what is in a food: sampling, proximate analysis, official methods, and the instruments — spectroscopy, chromatography, mass spectrometry and electrophoresis — that a food analyst runs. It is the core of the food analyst qualification in countries that license analysts.

Our 65 ICAR AIEEA-PG / PGS questions in this subject: 17 easy, 42 medium, 6 hard, across 14 units.

10 practice questions, worked

Choose an answer, then open the working. Each explanation says why the right answer is right and why the others are not.

These 10 were written to ICAR AIEEA-PG / PGS and are not shown on any other exam's page.

Question 1 · easy

Coning and quartering is used to

  1. reduce a granular sample representatively
  2. dissolve fat from a meat sample before weighing
  3. digest organic matter before mineral analysis
  4. separate proteins by molecular weight
Answer and working

A. reduce a granular sample representatively

The heaped sample is flattened, divided into quarters and opposite quarters kept, repeating until the size is suitable.

Question 2 · medium

For vitamin analysis, samples are best handled

  1. in bright sunlight to speed up extraction of the vitamins
  2. in subdued light, cold, with antioxidants
  3. at 105 °C to remove water before extraction begins
  4. in open containers so that oxygen can stabilise them
Answer and working

B. in subdued light, cold, with antioxidants

Many vitamins are destroyed by light, heat and oxygen; protective handling preserves them.

Question 3 · medium

Oven moisture methods are described as empirical because they

  1. give exactly the true water content of every food, whatever its volatiles
  2. use no balance at all
  3. depend only on the colour of the dried sample
  4. measure everything lost under the stated conditions
Answer and working

D. measure everything lost under the stated conditions

Volatile compounds and decomposition losses count; conditions define the result.

Question 4 · easy

A sample has 13.0 % moisture. Its total solids content is

  1. 13.0 %
  2. 113.0 %
  3. 7.7 %
  4. 87.0 %
Answer and working

D. 87.0 %

Total solids = 100 − moisture = 87.0 %.

Question 5 · hard

A jam has 2.1 % reducing sugars before inversion and 8.4 % after inversion (as invert sugar). Its sucrose content is

  1. 5.99 %
  2. 6.30 %
  3. 8.40 %
  4. 6.63 %
Answer and working

A. 5.99 %

Sucrose = (8.4 − 2.1) × 0.95 = 5.99 %. 6.30 % omits the 0.95 factor; 8.40 % is total after inversion; 6.63 % divides by 0.95.

Question 6 · medium

AOAC 985.29 (Prosky) determines total dietary fibre by

  1. boiling with dilute acid and alkali and weighing the dried residue
  2. HPLC of free sugars
  3. enzymatic digestion, ethanol precipitation and weighing
  4. near-infrared spectroscopy only
Answer and working

C. enzymatic digestion, ethanol precipitation and weighing

Amylase, protease and amyloglucosidase remove starch and protein; soluble fibre is precipitated with ethanol; residue is corrected for protein and ash.

Question 7 · hard

Two HPLC peaks elute at 6.2 and 7.0 min with baseline widths of 0.40 min each. Their resolution is

  1. 1.0
  2. 0.8
  3. 4.0
  4. 2.0
Answer and working

D. 2.0

Rs = 2(7.0 − 6.2)/(0.40 + 0.40) = 2.0, baseline separated. 1.0 omits the factor 2; 0.8 is the time difference; 4.0 divides by one width.

Question 8 · medium

In reversed-phase HPLC on a C18 column with water–methanol, the compounds that elute first are

  1. the most polar
  2. the most non-polar
  3. the largest
  4. the most volatile
Answer and working

A. the most polar

Polar compounds interact weakly with the non-polar stationary phase and elute early.

Question 9 · medium

A resolution of at least 1.5 between two peaks indicates

  1. complete co-elution
  2. baseline separation
  3. a detector fault
  4. the column is overloaded
Answer and working

B. baseline separation

Rs ≥ 1.5 means the peaks are separated to baseline.

Question 10 · medium

Sandwich ELISA is suited to food allergens because allergenic proteins

  1. are too small to bind any antibody
  2. have at least two antibody-binding sites
  3. are volatile and leave the sample
  4. fluoresce strongly
Answer and working

B. have at least two antibody-binding sites

Capture and detection antibodies bind different epitopes of the large protein.

A study plan for Food Analysis: Proximate, Chemical & Instrumental Methods in ICAR AIEEA-PG / PGS

This plan follows where our questions actually fall, so the time goes where the marks are likely to be. It is our reading of the syllabus, not the exam body's weighting.

  1. First, the heavy units (45% of the 65 questions): Protein: Kjeldahl and Dumas methods (7, 11%); Chromatography: TLC, GC and HPLC (6, 9%); Physical tests: pH, titratable acidity, refractometry and colour (6, 9%); Ash and mineral analysis (5, 8%); Carbohydrates, sugars and dietary fibre (5, 8%).

    Read the manual chapters for these first and do every worked question you can find on them.

  2. Then the middle band: Fat: Soxhlet, Mojonnier and Gerber methods (5, 8%); Moisture and total solids (5, 8%); Sampling and sample preparation (5, 8%); UV-visible, fluorescence and infrared spectroscopy (5, 8%); Electrophoresis and immunoassays: SDS-PAGE and ELISA (4, 6%).

    Aim for one sitting per unit, then mixed questions across them.

  3. Then the lighter units: Official methods and method compendia: AOAC, ISO and national manuals (4, 6%); Vitamin analysis (4, 6%); Atomic absorption and ICP spectroscopy (3, 5%); Mass spectrometry and hyphenated techniques (1, 2%).

    Learn the definitions, classifications and standard numbers; these units are more often recalled than worked.

  4. Leave time for the hard ones: Ash and mineral analysis (1 hard of 5), Carbohydrates, sugars and dietary fibre (1 hard of 5), UV-visible, fluorescence and infrared spectroscopy (1 hard of 5).

    Come back to these after a first pass, untimed, and read every explanation, including for the options you rejected.

  5. Bank the easy marks: Mass spectrometry and hyphenated techniques (1 easy of 1); Fat: Soxhlet, Mojonnier and Gerber methods (2 easy of 5).

    Do these early in revision so they are fast on the day.

  6. Finish timed. Sit the subject as a timed set without notes, mark it, and send each wrong answer back to its unit above.

Last updated · How Laisha writes and checks its pages · Report a mistake

Ask Laisha about this

Opens Laisha's chat with your question ready to send. Sign in or start free.

Unlock every question — ICAR AIEEA-PG / PGS Food Analysis: Proximate, Chemical & Instrumental Methods

US$10 a month · one price worldwide · cancel any time

  • Every practice question in this subject, with full working
  • Timed exam papers and progress by unit
  • Ask Laisha to explain any answer — up to 1,000 questions a month
Free or Smarter
FreeSmarter
Grow guides, pest pages, articlesRead allRead all
Ask Laisha5 a dayUp to 1,000 a month
Exam practiceThe sample on each pageEvery question, timed papers
Crop calendar and pest alerts—For your own fields
Calculators and tools—All
Uploads and reports—100 a month, real files
10 days free for every new account, no card.